UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
Ref. No.
SOP/PD/206_01
STANDARD OPERATING PROCEDURE
FORCE TEN
®
FDX
Valid for:
2 years from
approval
Page 2 of 11
1. Scope
This Standard Operating Procedure (SOP) applies to the staff and students
using the Force TEN
®
FDX in the Pharmacy Practice Resource Unit (PPRU)
at the Pharmacy Department, University of Malta.
2. Objective
To describe the procedure for the set-up, operation and calibration of the Force
TEN
®
FDX.
3. Definitions
3.1. Auto-Calibration: To return the device to the specified accuracy when
establishing that the gauge is out of tolerance after carrying out a
calibration verification test.
3.2. ON/OFF Key: To turn the device on or off.
3.3. Calibration Verification: A verification test carried out using certified
weights used to determine whether a calibration procedure for the
device is necessary.
3.4. ESCAPE Function: An additional function of the UNITS key used to
cancel and return to the previous menu option and to exit menu set-up.
3.5. Force TEN
®
FDX: A digital gauge designed for routine hand-held
force testing. It continually captures both peak tension and peak
compression forces. May also be used as an algometer to test pain
threshold, pain tolerance, trigger and tender points as well as to
monitor treatment progress.
3.6. Non-Peak Mode: The device will display the instantaneous force
being applied. Used to monitor the applied force as it is being varied.
3.7. PEAK Key: To activate peak mode and recall peak readings.
3.8. Peak Memory: The saved peak readings.
3.9. Peak Mode: The device will display the highest peak tension or
compression force being applied since the last time peak memory was