UNIVERSITY OF MALTA
FACULTY OF MEDICINE & SURGERY
PHARMACY DEPARTMENT
Ref. No.
SOP/PD/206_01
STANDARD OPERATING PROCEDURE
FORCE TEN
®
FDX
Valid for:
2 years from
approval
Page 5 of 11
5.3.1.4. Apply each weight consecutively in both tension and
compression.
5.3.1.5. If the displayed weight differs by > +/- 0.3% of the
full scale +/-1 least significant figure, the device is
out of tolerance and requires auto-calibration (see
section 5.3.2).
5.3.2. Auto-Calibration
5.3.2.1. Repeatedly press the UNITS key until the arrow is
pointing towards lbf since all calibration procedures
are to be performed in pounds.
5.3.2.2. Keep the device in a still position.
5.3.2.3. Enter the Set-Up Menu (see 5.1.3 to 5.1.7).
5.3.2.4. Repeatedly press the SCROLL key until [L CAL] is
displayed.
5.3.2.5. Press the SELECT key twice to display the force cell
capacity.
5.3.2.6. If the displayed force cell capacity does not match
with that of the force cell module, repeatedly press
the SCROLL key until the correct force cell capacity
is displayed.
5.3.2.7. Apply the test weight that is equal to the displayed
full-scale capacity weight.
5.3.2.8. Press the SELECT key.
5.3.2.9. Wait until the device displays the applied weight to
indicate that the calibration process was successful.
5.3.2.10. If [uuuuu] “under” or [nnnnn] “over” is displayed,
the weight used was not equal to the full-scale
capacity of the force cell module and auto-
calibration must be repeated with an appropriate
standard weight.